【摘要】 目的 研究兩種不同的衰減系數轉換方法對正電子發射計算機斷層顯像/計算機體層掃描(PET/CT) 標準攝取值(SUV)測量值的影響。 方法 2009年11-12月,從經過PET/CT氟代脫氧葡萄糖(FDG)顯像患者中選取20例,其口腔中均有金屬假牙。以四段轉換法和二段轉換法重建全身衰減校正斷層圖像。分別選取3種高密度區域和7種低密度區域,測量其最大SUV和平均SUV,比較兩種轉換方法SUV測量值。 結果 平均SUV:二段轉換法的3種高密度區域SUV降低,在7種低密度區域中,1種區域增高,1種區域降低(P值均 lt;0.05)。最大SUV:二段轉換法的2種高密度區域的SUV降低,在7種低密度區域中,1種區域增高,2種靠近高密度組織的區域降低(P值均 lt;0.05)。 結論 二段轉換法能降低高密度區域的SUV,可用于減小體內金屬植入物和CT對比劑造成的過度校正。
【Abstract】 Objective To explore the effect of two attenuation correction algorithms on PET/CT SUV measurement. Methods From November to December 2009, the PET Slice of 20 patients with metallic dental implant were reconstructed with four-and two-section algorithms respectively. Mean SUV and maximum SUV were measured in three high-density areas and 7 low-density areas. Paired t test were performed to compare the differences. Results Mean SUV: two-section algorithm produced significantly lower SUV in all the three high-density areas; in the 7 low-density areas, SUV increased obviously in one area and decreased apparently in one area (P lt;0.05). Maximum SUV: two-section algorithm produced significantly lower SUV in two high-density areas, SUV increased obviously in one area and decreased apparently in two areas which was adjacent to the high density areas (P lt;0.05). Conclusion Two-section algorithm produces lower SUV measurement value than the four-section algorithm does, and it is useful in PET/CT studies for patients with metallic dental implant and when CT contrast is used.
引用本文: 楊曉川,周綠漪. PET/CT衰減系數轉換方法對標準攝取值測量的影響研究. 華西醫學, 2010, 25(11): 2052-2054. doi: 復制