• 1 四川大學華西醫院實驗醫學科臨檢血液室(成都,610041);2 資陽市第一人民醫院;
導出 下載 收藏 掃碼 引用

【摘要】 目的  探討全自動尿沉渣分析儀在尿管型檢測中的應用。 方法  收集尿管型患者晨尿標本836份。所有標本均經尿干化學分析儀檢測Pro≥1+或尿沉渣分析儀提示有管型。采用UF100全自動尿沉渣分析儀和顯微鏡檢測管型,對比分析兩者的檢測結果。 結果  836份標本中,UF100全自動尿沉渣分析儀檢測陽性者320例,占38.28%;顯微鏡檢測陽性者195例,占23.33%。其中UF100全自動尿沉渣分析儀的假陽性率為26.52%,假陰性率為23.08%。UF100尿沉渣分析儀與顯微鏡檢測管型的陽性結果比較,差異有統計學意義(P lt;0.01)。 結論  UF100全自動尿沉渣儀能快速篩檢尿沉渣,但存在一定的假陰性,必須同時將其檢測結果與尿干化學結果結合考慮以決定是否再進行顯微鏡檢測,減少假陰性以防止漏檢。
【Abstract】 Objective  To investigate the application of UF-100 full-auto urine sediment analyzer in detecting cylindruia. Methods  836 specimens with cylindruia were selected. All the specimens with Pro≥1+ were dectected by chemical dipstick or cylindruia by urine sediment analyzer. The cast were detected by urine sediment analyzer and microscope, and the results were compared. Results  Of 836 specimens, 320 positive samples(38.28%) were found by UF-100 while 195 (23.33%) were found by microscope. False positive rate and false negative rate of UF100 were 26.52% and 23.08%. Compared the results of urine sediment analyzer with microscope, the difference was statistically significant (P lt;0.01). Conclusions  UF100 can detect urinary cast quickly, but there is a little high false negative rate. So we should consider urine sample whether to be detected by microscope compared with results of UF100 and chemical dipstick.

引用本文: 鄧山鷹,林勝. 全自動尿沉渣分析儀在尿管型檢測中的應用. 華西醫學, 2010, 25(8): 1515-1516. doi: 復制