• 1. Eye Institute and Affiliated Xiamen Eye Center of Xiamen University, Xiamen Clinical Research Center for Eye Diseases, Xiamen Key Laboratory of Ophthalmology, Fujian Provincial Key Laboratory of Corneal & Ocular Surface Diseases, Xiamen Municipal Key Laboratory of Corneal & Ocular Surface Diseases, Translational Medicine Institute of Xiamen Eye Center of Xiamen University, Xiamen 361001, China;
  • 2. Department of Ophthalmology, Peking University People’s Hospital, Beijing 100044, China;
  • 3. Department of Ophthalmology, Peking University International Hospital, Beijing 102206, China;
Li Xiaoxin, Email: drlixiaoxin@163.com
Export PDF Favorites Scan Get Citation

Objective To evaluate the performance, including stability and cutting efficiency, of a domestically produced S model dual-blade, dual-port vitrectomy probe during vitrectomy surgery. Methods A prospective, randomized controlled, non inferiority, multicenter clinical trial study. 120 patients who underwent PPV treatment at Xiamen Eye Center, Department of Ophthalmology of Peking University People’s Hospital, and Department of Ophthalmology, Peking University International Hospital in August 2025 were included in the study. According to the cutting frequency and instrument source, they are divided into 4 groups: 10 k cpm domestic S-type vitrectomy probe group, imported A-type vitrectomy probe group, and 20 k cpm S-type vitrectomy probe group and A-type vitrectomy probe group, with 30 cases in each group. The S-type vitrectomy probe adopts a double-edged dual air path design at frequencies of 10 and 20 k cpm; The A-type vitrectomy probe has a single blade design at a frequency of 10 k cpm and a double-edged design at a frequency of 20 k cpm. Compare and observe the differences in cutting efficiency, safety (incidence of iatrogenic retinal tears), and equipment stability (bubble generation, frequency and negative pressure changes, defect detection) between two groups. The comparison between groups was conducted using the chi square test. Results Due to missing data from one patient in the 20 k cpm S-type vitrectomy probe group, 119 cases were ultimately included in the statistical analysis. Among them, there were 59 cases in the S-type probe group (30 cases at 10 k cpm and 29 cases at 20 k cpm), and 60 cases in the A-type probe group (30 cases at 10 k cpm and 20 k cpm each). There was no significant difference in age, sex composition ratio, eye type, whether to combine cataract surgery, anesthesia mode, disease classification, proliferative vitreoretinopathy and proliferative membrane grading of proliferative diabetes retinopathy between 10 k cpm, 20 k cpm S-type vitrectomy probe group and A-type vitrectomy probe group (P>0.05). There was no statistically significant difference in cutting efficiency, cutting safety, equipment stability, frequency and negative pressure fluctuation amplitude, and defect detection between the two groups (P>0.05). In terms of bubble generation in equipment stability, the 20 k cpm S-type vitrectomy probe group performed significantly better than the A-type vitrectomy probe group, and the difference was statistically significant (P=0.027). Conclusion The domestic S-type vitrectomy probe is non-inferior to the imported A-type vitrectomy probe across multiple performance indicators and shows superior anti-bubble performance.

Citation: Li Yang, Miao Heng, Yuan Mengke, Li Xiaoxin. A study on the stability and cutting efficiency of a domestically produced dual-blade, dual-port vitrectomy probe. Chinese Journal of Ocular Fundus Diseases, 2026, 42(10): 842-848. doi: 10.3760/cma.j.cn511434-20251201-00543 Copy

Copyright ? the editorial department of Chinese Journal of Ocular Fundus Diseases of West China Medical Publisher. All rights reserved

  • Previous Article

    Interpretation of Expert consensus on quality control for intravitreal injections in China (2026)
  • Next Article

    Clinical observation of endoscopic assisted vitrectomy for the patients with eye atrophy